Skip to Navigation | Skip to Content

Senior Design 03-04: Team 9


Alternative Testing Methods

Hydrostatic Pressure Decay

Hydrostatic leak detection is conducted by pressurizing the test material from the inside with a liquid. The component to be tested is filled completely with a liquid, usually water. The liquid is then pressurized to a required testing pressure. This pressure is then maintained for a set time at which point the component can be visually inspected to locate leaks. Often times, a die or tracer is added to the testing liquid to make it easier to visually locate the leak.

This ability to visually locate the leak is a big advantage over other leak testing methods, which only test on a “go-nogo” system. That is, if the purpose of the test is to determine not only if the product leaks, but also where it leaks, then the hydrostatic method should be strongly considered as the method of choice.

Other detection methods can also locate the leak, but are often more costly and prone to error. For instance, helium testing is a very good method to locate leaks, but the equipment can be very expensive depending on how small of a leak needs to be detected. Helium leak detectors are also very prone to seal failure. Often times, the seals need to be soldered onto the material in order to ensure that no helium molecules escape around the seal. The hydrostatic method is much more appropriate on leaks that are not on the molecular order, and therefore do not require such expensive equipment or molecular tight seals.

Disadvantages to hydrostatic leak testing would include necessary dry time after testing. Dry time can be a very, very costly thing depending on part cycle times as well as other testing and shipping requirements. Often times, eliminating the dry time can lead to much faster cycle times and accordingly a decrease in overall costs.

Helium Leak Testing

Helium Leak testing is one method for leak testing any closed circuit. One of the major uses for helium leak tests is for industrial pipelines, but it is also used for testing heat exchangers. The test for any component is basically the same. The chamber is sealed at the ends, helium is pumped in, a helium sensitive meter is then passed over, checking for leaks. The main use for leak tests is for localizing the leak, although flow rate can also be measured.

For conducting a pipeline leak test, the ends of the desired pipe section are sealed off. Helium is then injected into the pipeline and a technician scans the ground above the pipeline looking for any helium escaping from a leak. The reason for pipeline leak tests is for localization of leaks. For scanning the ground above, the technician uses a helium sniffer, which is a handheld device using a mass spectrometer that is sensitive to helium.

Another use for helium leak testing is for Heat exchangers. For heat exchangers, a vacuum helium leak test is done. In doing this type of test, the machine will first carry out a gross leak test using air on the component, so the helium is not wasted. Then the actual and accurate test will be carried out using helium in the chamber. The helium wand/sniffer is passed around the chamber by the technician and any quantity of helium that permeates through is picked up by the mass spectrometer.
Vacuum leak testing is a faster test than a pressure type based test. Reducing manufacturer cycle time. Both leak location identification and leak rate measurements are possible. Flow may be liquid, gaseous or both. Helium can be measured at a level below that which is present in the atmosphere.
Some major issues concerning helium leak testing include the following:

  1. The location of the sniffer probe is operator dependent. The reading depends on the duration, angle, and distance of the probe away from the leak.
  2. Air currents are always present and affect the air to helium ratio that can be detected.
  3. The type of permeation of the leak will affect the flow rate.
  4. Helium is present in the atmosphere and will affect the sensitivity of the meter.